StructureDefinition-PhdBaseObservation

Sourcehl7.fhir.uv.phd#current:Personal Health Device Implementation Guide (v4.0.1)
resourceTypeStructureDefinition
idPhdBaseObservation
canonicalhttp://hl7.org/fhir/uv/phd/StructureDefinition/PhdBaseObservation
version1.0.0
statusdraft
publisherHL7 International - Health Care Devices Work Group
namePhdBaseObservation
date2017-11-24T15:17:35.385-05:00
descriptionObservation elements that are common to all the Continua PHD Observation profiles
jurisdictionsuv
fhirVersion4.0.1
kindresource
abstractfalse
sdTtypeObservation
derivationconstraint
basehttp://hl7.org/fhir/StructureDefinition/Observation
Usages
Name Flags Card. Type Description & Constraints doco
. . Observation Observation
. . . meta 1..
. . . . Slices for security Slice: Unordered, Open by value:system, value:code
. . . . . security:PhdHtest ..1 Security value set to HTEST when measurement status indicates demo or test data
. . . . . . system 1.. Fixed Value: http://terminology.hl7.org/CodeSystem/v3-ActReason
. . . . . . code 1.. Fixed Value: HTEST
. . . extension 0..1 Extension for the PHG
. . . . url "http://hl7.org/fhir/StructureDefinition/observation-gatewayDevice"
. . . . valueReference 1.. Reference () Reference to the Personal Health Gateway (PHG) Device
. . . Slices for identifier Slice: Unordered, Open by exists:value, exists:system, exists:type
. . . . identifier:conditionalCreate ..1 Unique identifier of this measurement for a given patient and device
. . . . . type .. 0
. . . . . system .. 0
. . . . . value 1..
. . . status
. . . code
. . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . coding:MDCType 1..1 The 11073-10101 MDC code for the measurement
. . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . code 1..
. . . . . coding:LoincCoding ..1 The LOINC code for the measurement
. . . . . . system 1.. Fixed Value: http://loinc.org
. . . . . . code 1..
. . . . text It is recommended to display at least the reference identifier for the MDC code
. . . subject 1..
. . . . reference 1..
. . . effective[x] 1..
. . . . effectiveDateTime dateTime
. . . . effectivePeriod Period
. . . interpretation
. . . . Slices for coding Slice: Unordered, Open by value:system
. . . . . coding:MsmtStatusInterpretation 1..1 Measurement status interpretation when questionable, calibration-ongoing, validated, early-indication, threshold exceeded, alarm inhibited
. . . . . . system 1.. Fixed Value: http://hl7.org/fhir/uv/pocd/CodeSystem/measurement-status
. . . . . . code 1..
. . . device 1..
. . . . reference 1.. Reference to the device responsible for the measurement
. . . derivedFrom Reference to the Coincident Time Stamp Observation and/or Source-Handle-Reference Observation
. . . Slices for component Slice: Unordered, Open by value:code
. . . . component:supplementalTypesComponent Supplemental Type: A further description of the measurement type.
. . . . . code Binding: CodeableConcept 11073 Attributes ( required )
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:MdcType 1..1 The 11073-10101 MDC code for the measurement
. . . . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . . . code 1.. Fixed Value: 68193
. . . . . . text
. . . . . value[x]:valueCodeableConcept 1.. CodeableConcept
. . . . . valueCodeableConcept
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:MDCType 1..1 Required MDC code entry.
. . . . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . . . code 1..
. . . . . dataAbsentReason .. 0
. . . . component:relativeTimeComponent ..1 Relative time stamp of the measurement reported by the PHD.
. . . . . code Binding: Quantity 11073 Attributes ( required )
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:MdcType 1..1 The 11073-10101 MDC code for the measurement
. . . . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . . . code 1.. Fixed Value: 67985
. . . . . . text
. . . . . value[x]:valueQuantity 1.. Quantity
. . . . . valueQuantity
. . . . . . value 1..
. . . . . . system 1.. Fixed Value: http://unitsofmeasure.org
. . . . . . code 1.. Fixed Value: us
. . . . . dataAbsentReason .. 0
. . . . component:hiresRelativeTimeComponent ..1 Hi-Resolution Relative time stamp of the measurement reported by the PHD.
. . . . . code Binding: Quantity 11073 Attributes ( required )
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:MdcType 1..1 The 11073-10101 MDC code for the measurement
. . . . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . . . code 1.. Fixed Value: 68073
. . . . . . text
. . . . . value[x]:valueQuantity 1.. Quantity
. . . . . valueQuantity
. . . . . . value 1.. Numerical value
. . . . . . system 1.. Fixed Value: http://unitsofmeasure.org
. . . . . . code 1.. Fixed Value: us
. . . . . dataAbsentReason .. 0

doco Documentation for this format

Produced 06 Apr 2023