StructureDefinition-PhdCompoundNumericObservation

Sourcehl7.fhir.uv.phd#current:Personal Health Device Implementation Guide (v4.0.1)
resourceTypeStructureDefinition
idPhdCompoundNumericObservation
canonicalhttp://hl7.org/fhir/uv/phd/StructureDefinition/PhdCompoundNumericObservation
version1.0.0
statusdraft
publisherHL7 International - Health Care Devices Work Group
namePhdCompoundNumericObservation
date2017-06-02T14:29:52.39367-04:00
descriptionStructureDefinition for Observation Resources representing measurement data coming from a medical device where the measurement is described by a set of numbers such as the x, y, and x components of an acceleration.
jurisdictionsuv
fhirVersion4.0.1
kindresource
abstractfalse
sdTtypeObservation
derivationconstraint
basehttp://hl7.org/fhir/uv/phd/StructureDefinition/PhdBaseObservation
Usages(none)
Name Flags Card. Type Description & Constraints doco
. . Observation PhdBaseObservation
. . . meta 1..
. . . . Slices for profile 1 .. Slice: Unordered, Open by value:value
. . . . . profile:phdProfile 1..1 Fixed Value: http://hl7.org/fhir/uv/phd/StructureDefinition/PhdCompoundNumericObservation
. . . category:vitalSignsCategory Required by the FHIR specification if the measurement value is a vital sign
. . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . coding:vitalSigns 1..1 Required Vital Signs category code system entry
. . . . . . system 1.. Fixed Value: http://terminology.hl7.org/CodeSystem/observation-category
. . . . . . code 1.. Fixed Value: vital-signs
. . . value[x] .. 0
. . . dataAbsentReason
. . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . coding:FhirDefault 1..1
. . . . . . system 1.. Fixed Value: http://terminology.hl7.org/CodeSystem/data-absent-reason
. . . . . . code 1..
. . . Slices for component Slice: Unordered, Open by value:code
. . . . component:compound Compound numeric measurement entry components
. . . . . code Binding: 11073 MDC Non-Object codes ( required )
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:MDCType 1..1 The 11073-10101 MDC code for the measurement
. . . . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . . . code 1..
. . . . . . . coding:LoincCoding ..1 The LOINC code for the measurement if a vital sign
. . . . . . . . system 1.. Fixed Value: http://loinc.org
. . . . . . . . code 1..
. . . . . . text
. . . . . value[x] Quantity
. . . . . . value 1..
. . . . . . system 1.. Fixed Value: http://unitsofmeasure.org
. . . . . . code 1..
. . . . . dataAbsentReason Populated when a special value or status in Compound-Nu-Observed-Value indicates invalid, not unavailable, or msmt ongoing
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:FhirDefault 1..1
. . . . . . . . system 1.. Fixed Value: http://terminology.hl7.org/CodeSystem/data-absent-reason
. . . . . . . . code 1..
. . . . . interpretation
. . . . . . Slices for coding Slice: Unordered, Open by value:system
. . . . . . . coding:MsmtStatusInterpretation 1..1 Measurement status interpretation when questionable, calibration-ongoing, validated, early-indication, threshold exceeded, alarm inhibited
. . . . . . . . system 1.. Fixed Value: http://hl7.org/fhir/uv/pocd/CodeSystem/measurement-status
. . . . . . . . code 1..
. . . . component:accuracyComponent ..1 The accuracy of the measurement
. . . . . code Binding: Quantity 11073 Attributes ( required )
. . . . . . Slices for coding 1 .. Slice: Unordered, Open by value:system
. . . . . . . coding:MdcType 1..1 The 11073-10101 MDC code for the measurement
. . . . . . . . system 1.. Fixed Value: urn:iso:std:iso:11073:10101
. . . . . . . . code 1.. Fixed Value: 67914
. . . . . . text
. . . . . value[x]:valueQuantity 1.. Quantity
. . . . . . value 1..
. . . . . . system 1.. Fixed Value: http://unitsofmeasure.org
. . . . . . code 1..
. . . . . dataAbsentReason .. 0