Name
|
Flags
|
Card.
|
Type
|
Description & Constraints
|
Observation
|
|
|
PhdBaseObservation
|
|
meta
|
|
1.. |
|
|
Slices for profile
|
|
1
..
|
|
Slice:
Unordered, Open by value:value
|
profile:phdProfile
|
|
1..1 |
|
Fixed Value:
http://hl7.org/fhir/uv/phd/StructureDefinition/PhdCompoundNumericObservation
|
category:vitalSignsCategory
|
|
|
|
Required by the FHIR specification if the measurement value is a vital sign |
Slices for coding
|
|
1
..
|
|
Slice:
Unordered, Open by value:system
|
coding:vitalSigns
|
|
1..1 |
|
Required Vital Signs category code system entry |
system
|
|
1.. |
|
Fixed Value:
http://terminology.hl7.org/CodeSystem/observation-category
|
code
|
|
1.. |
|
Fixed Value:
vital-signs
|
value[x]
|
|
..
0
|
|
|
dataAbsentReason
|
|
|
|
|
Slices for coding
|
|
1
..
|
|
Slice:
Unordered, Open by value:system
|
coding:FhirDefault
|
|
1..1 |
|
|
system
|
|
1.. |
|
Fixed Value:
http://terminology.hl7.org/CodeSystem/data-absent-reason
|
code
|
|
1.. |
|
|
Slices for component
|
|
|
|
Slice:
Unordered, Open by value:code
|
component:compound
|
|
|
|
Compound numeric measurement entry components |
code
|
|
|
|
Binding:
11073 MDC Non-Object codes
( required
)
|
Slices for coding
|
|
1
..
|
|
Slice:
Unordered, Open by value:system
|
coding:MDCType
|
|
1..1 |
|
The 11073-10101 MDC code for the measurement |
system
|
|
1.. |
|
Fixed Value:
urn:iso:std:iso:11073:10101
|
code
|
|
1.. |
|
|
coding:LoincCoding
|
|
..1 |
|
The LOINC code for the measurement if a vital sign |
system
|
|
1.. |
|
Fixed Value:
http://loinc.org
|
code
|
|
1.. |
|
|
text
|
|
|
|
|
value[x]
|
|
|
Quantity |
|
value
|
|
1.. |
|
|
system
|
|
1.. |
|
Fixed Value:
http://unitsofmeasure.org
|
code
|
|
1.. |
|
|
dataAbsentReason
|
|
|
|
Populated when a special value or status in Compound-Nu-Observed-Value indicates invalid, not unavailable, or msmt ongoing |
Slices for coding
|
|
1
..
|
|
Slice:
Unordered, Open by value:system
|
coding:FhirDefault
|
|
1..1 |
|
|
system
|
|
1.. |
|
Fixed Value:
http://terminology.hl7.org/CodeSystem/data-absent-reason
|
code
|
|
1.. |
|
|
interpretation
|
|
|
|
|
Slices for coding
|
|
|
|
Slice:
Unordered, Open by value:system
|
coding:MsmtStatusInterpretation
|
|
1..1 |
|
Measurement status interpretation when questionable, calibration-ongoing, validated, early-indication, threshold exceeded, alarm inhibited |
system
|
|
1.. |
|
Fixed Value:
http://hl7.org/fhir/uv/pocd/CodeSystem/measurement-status
|
code
|
|
1.. |
|
|
component:accuracyComponent
|
|
..1 |
|
The accuracy of the measurement |
code
|
|
|
|
Binding:
Quantity 11073 Attributes
( required
)
|
Slices for coding
|
|
1
..
|
|
Slice:
Unordered, Open by value:system
|
coding:MdcType
|
|
1..1 |
|
The 11073-10101 MDC code for the measurement |
system
|
|
1.. |
|
Fixed Value:
urn:iso:std:iso:11073:10101
|
code
|
|
1.. |
|
Fixed Value:
67914
|
text
|
|
|
|
|
value[x]:valueQuantity
|
|
1.. |
Quantity |
|
value
|
|
1.. |
|
|
system
|
|
1.. |
|
Fixed Value:
http://unitsofmeasure.org
|
code
|
|
1.. |
|
|
dataAbsentReason
|
|
..
0
|
|
|
Documentation for this format
|