Source | hl7.fhir.uv.phd#current:Personal Health Device Implementation Guide (v4.0.1) |
resourceType | StructureDefinition |
id | PhdCodedEnumerationObservation |
canonical | http://hl7.org/fhir/uv/phd/StructureDefinition/PhdCodedEnumerationObservation |
version | 1.0.0 |
status | draft |
publisher | HL7 International - Health Care Devices Work Group |
name | PhdCodedEnumerationObservation |
date | 2017-06-02T14:29:52.39367-04:00 |
description | StructureDefinition for Observation Resources representing measurement data coming from a medical device where the measurement itself is an enumerated code. |
jurisdictions | uv |
fhirVersion | 4.0.1 |
kind | resource |
abstract | false |
sdTtype | Observation |
derivation | constraint |
base | http://hl7.org/fhir/uv/phd/StructureDefinition/PhdBaseObservation |
Usages | (none) |
Name | Flags | Card. | Type | Description & Constraints |
---|---|---|---|---|
Observation | PhdBaseObservation | Reports measurements that are enumeration MDC codes | ||
meta | 1.. | |||
Slices for profile | 1 .. | Slice: Unordered, Open by value:value | ||
profile:phdProfile | 1..1 | Fixed Value: http://hl7.org/fhir/uv/phd/StructureDefinition/PhdCodedEnumerationObservation | ||
valueCodeableConcept | CodeableConcept | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MDCType | 1..1 | Required MDC code entry | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
dataAbsentReason | Populated when the Measurement-Status or an Enum-Observed-Value status indicates invalid, not unavailable, or msmt ongoing | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:FhirDefault | 1..1 | |||
system | 1.. | Fixed Value: http://terminology.hl7.org/CodeSystem/data-absent-reason | ||
code | 1.. | |||
Documentation for this format |