Source | hl7.fhir.uv.phd#current:Personal Health Device Implementation Guide (v4.0.1) |
resourceType | StructureDefinition |
id | PhdNumericObservation |
canonical | http://hl7.org/fhir/uv/phd/StructureDefinition/PhdNumericObservation |
version | 1.0.0 |
status | draft |
publisher | HL7 International - Health Care Devices Work Group |
name | PhdNumericObservation |
date | 2017-06-02T14:29:52.39367-04:00 |
description | StructureDefinition for Observation Resources representing measurement data coming from a medical device where the measurement is number. |
jurisdictions | uv |
fhirVersion | 4.0.1 |
kind | resource |
abstract | false |
sdTtype | Observation |
derivation | constraint |
base | http://hl7.org/fhir/uv/phd/StructureDefinition/PhdBaseObservation |
Usages | (none) |
Name | Flags | Card. | Type | Description & Constraints |
---|---|---|---|---|
Observation | PhdBaseObservation | Simple Numeric measurement | ||
meta | 1.. | |||
Slices for profile | 1 .. | Slice: Unordered, Open by value:value | ||
profile:phdProfile | 1..1 | Fixed Value: http://hl7.org/fhir/uv/phd/StructureDefinition/PhdNumericObservation | ||
category:vitalSignsCategory | Required by the FHIR specification if the measurement value is a vital sign | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:vitalSigns | 1..1 | Required Vital Signs category code system entry | ||
system | 1.. | Fixed Value: http://terminology.hl7.org/CodeSystem/observation-category | ||
code | 1.. | Fixed Value: vital-signs | ||
value[x] | Quantity | |||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | The UCUM code for the units of this measurement. | ||
dataAbsentReason | Populated when a special value occurs or Measurement-Status indicates invalid, not unavailable, or msmt ongoing | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:FhirDefault | 1..1 | |||
system | 1.. | Fixed Value: http://terminology.hl7.org/CodeSystem/data-absent-reason | ||
code | 1.. | |||
component:accuracyComponent | ..1 | The accuracy of the measurement | ||
code | Binding: Quantity 11073 Attributes ( required ) | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MdcType | 1..1 | The 11073-10101 MDC code for the measurement | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | Fixed Value: 67914 | ||
text | ||||
value[x]:valueQuantity | 1.. | Quantity | ||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | |||
dataAbsentReason | .. 0 | |||
component:alertOpStateComponent | The Alert Operations State component | |||
code | Binding: ANS1ToHL7 codes defined for non-measurement attributes ( required ) | |||
coding | 1..1 | Only the ASN1 to HL7 coding for this item. | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | ||
code | 1.. | |||
value[x] | 1.. | CodeableConcept | ||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:V2Binary | 1..1 | The required Version 2 Table 0136 Binary code entry | ||
system | 1.. | Fixed Value: http://terminology.hl7.org/CodeSystem/v2-0203 | ||
code | 1.. | |||
dataAbsentReason | .. 0 | |||
component:currentLimitsComponent | ..1 | The Current-Limits values | ||
code | Binding: Range 11073 Attributes ( required ) | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MdcType | 1..1 | The 11073-10101 MDC code for the measurement | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | Fixed Value: 67892 | ||
text | ||||
value[x] | 1.. | Range | ||
low:SimpleQuantity | 1.. | |||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | |||
high:SimpleQuantity | 1.. | |||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | |||
dataAbsentReason | .. 0 | |||
component:alertOpTextStringComponent | ..1 | The Alert-Op-Text human readable strings | ||
code | Binding: String 11073 Attributes ( required ) | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MdcType | 1..1 | The 11073-10101 MDC code for the measurement | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | Fixed Value: 68104 | ||
text | ||||
value[x]:valueString | 1.. | string | ||
value | 1.. | |||
dataAbsentReason | .. 0 | |||
component:measurementConfidence95Component | ..1 | The Measurement-Confidence-95 range | ||
code | Binding: Range 11073 Attributes ( required ) | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MdcType | 1..1 | The 11073-10101 MDC code for the measurement | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | Fixed Value: 68236 | ||
text | ||||
value[x] | 1.. | Range | ||
low:SimpleQuantity | 1.. | |||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | |||
high:SimpleQuantity | 1.. | |||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | |||
dataAbsentReason | .. 0 | |||
component:thresholdNotificationTextStringComponent | ..1 | The Threshold-Notification-Text-String attribute value entry | ||
code | Binding: String 11073 Attributes ( required ) | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MdcType | 1..1 | The 11073-10101 MDC code for the measurement | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | Fixed Value: 68232 | ||
text | ||||
value[x]:valueString | 1.. | string | ||
value | 1.. | |||
dataAbsentReason | .. 0 | |||
Documentation for this format |
Produced 06 Apr 2023