Code
|
Display
|
Definition
|
113860
|
15cm from Isocenter toward Source |
15cm from the isocenter towards the X-Ray source; See IEC 60601-2-43. |
113862
|
1cm above Tabletop |
1cm above the patient tabletop or cradle; See FDA Federal Performance Standard for Diagnostic X-Ray Systems §1020.32(d) (3). |
113861
|
30cm in Front of Image Input Surface |
30cm in front (towards the tube) of the input surface of the image receptor; See FDA Federal Performance Standard for Diagnostic X-Ray Systems §1020.32(d) (3). |
113864
|
15cm from Table Centerline |
15cm from the centerline of the X-Ray table and in the direction of the X-Ray source; See FDA Federal Performance Standard for Diagnostic X-Ray Systems §1020.32(d) (3). |
113941
|
In Detector Plane |
A segmented region of the detector surface within the irradiated area (but might not be near the center of the detector). |
113863
|
30cm above Tabletop |
30cm above the patient tabletop of cradle; See FDA Federal Performance Standard for Diagnostic X-Ray Systems §1020.32(d) (3). |
113965
|
100cm from X-Ray Source |
100cm from the X-Ray source along the central ray of the X-Ray beam. |
113865
|
Entrance exposure to a 4.2 cm breast thickness |
Standard breast means a 4.2 centimeter (cm) thick compressed breast consisting of 50 percent glandular and 50 percent adipose tissue.See Department of Health and Human Services, Food and Drug Administration. Mammography quality standards; final rule. Federal Register. Oct. 28, 1997; 68(208):55852-55994. |
113964
|
At Surface of Patient |
A point at the surface of the patient within the irradiated area where the X-Ray beam enters the patient (i.e. towards the tube). |