Source | hl7.fhir.uv.phd#current:Personal Health Device Implementation Guide (v4.0.1) |
resourceType | StructureDefinition |
id | PhgDevice |
canonical | http://hl7.org/fhir/uv/phd/StructureDefinition/PhgDevice |
version | 1.0.0 |
status | draft |
publisher | HL7 International - Health Care Devices Work Group |
name | PhgDevice |
date | 2017-07-07T11:39:51.3383228-04:00 |
description | Base StructureDefinition for the Device Resource for a PHG |
jurisdictions | uv |
fhirVersion | 4.0.1 |
kind | resource |
abstract | false |
sdTtype | Device |
derivation | constraint |
base | http://hl7.org/fhir/StructureDefinition/Device |
Usages | (none) |
Name | Flags | Card. | Type | Description & Constraints |
---|---|---|---|---|
Device | Device | |||
meta | 1.. | |||
Slices for profile | 1 .. | Slice: Unordered, Open by value:value | ||
profile:phdProfile | 1..1 | Fixed Value: http://hl7.org/fhir/uv/phd/StructureDefinition/PhgDevice | ||
Slices for identifier | 1 .. |
Information that uniquely describes the personal health device
Slice: Unordered, Open by value:type.coding.system, value:type.coding.code |
||
identifier:systemIdIdentifier | 1..1 | System Id identifier | ||
type | 1.. | |||
coding | 1..1 | Indicates this is the IEEE 11073 System Id identifier | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/uv/phd/CodeSystem/ContinuaDeviceIdentifiers | ||
code | 1.. | Fixed Value: SYSID | ||
system | 1.. |
EUI-64 system identifier Fixed Value: urn:oid:1.2.840.10004.1.1.1.0.0.1.0.0.1.2680 |
||
value | 1.. | System id value from System-Id attribute | ||
identifier:btmacAddressIdentifier | ..1 | Bluetooth MAC Transport address identifier | ||
type | 1.. | |||
coding | 1..1 | Indicates this is the Bluetooth Mac address identifier | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/uv/phd/CodeSystem/ContinuaDeviceIdentifiers | ||
code | 1.. | Fixed Value: BTMAC | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/sid/eui-48/bluetooth | ||
value | 1.. | |||
identifier:macAddressIdentifier | ..1 | MAC Transport address identifier | ||
type | 1.. | |||
coding | 1..1 | Indicates this is the Mac address identifier | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/uv/phd/CodeSystem/ContinuaDeviceIdentifiers | ||
code | 1.. | Fixed Value: ETHMAC | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/sid/eui-48/ethernet | ||
value | 1.. | |||
manufacturer | Manufacturer name | |||
serialNumber | Serial number | |||
modelNumber | Model number | |||
partNumber | ||||
type | 1.. | Indicates the device is a PHG. | ||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MDCType | 1..1 | Required MDC code system entry | ||
system | 1.. |
Identifies IEEE 11073 10101 coding system Fixed Value: urn:iso:std:iso:11073:10101 |
||
code | 1.. |
Indicates PHG Fixed Value: 531981 |
||
text | ||||
specialization | This element contains an entry for each supported specialization or a generic representation | |||
systemType | The specialization standard supported by the PHG. | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MDCType | 1..1 | Required MDC code system entry for specializations | ||
system | 1.. |
The 11073-10101 system identifier Fixed Value: urn:iso:std:iso:11073:10101 |
||
code | 1.. | The 11073-10101 specialization code. Generic indicates it supports all. | ||
version | The version of the specialization standard supported by the PHG | |||
version | 1.. | A firmware, hardware, software, internal protocol may be reported for a PHG. A Continua version shall be reported. | ||
type | 1.. | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:MDCType | 1..1 | Required MDC code system entry | ||
system | 1.. |
Indicates the codes come from the MDC coding system Fixed Value: urn:iso:std:iso:11073:10101 |
||
code | 1.. | A code indicating the type of version the Device.version.value refers to. | ||
value | 1.. | The version | ||
Slices for property | 1 .. | Slice: Unordered, Open by value:type | ||
property:bitProperties | Properties reported in BITs fields | |||
type |
Tells what the BITs item is Binding: ANS1ToHL7 codes defined for non-measurement attributes ( required ) |
|||
coding | 1..1 | Required ASN1 code system entry | ||
system | 1.. | Fixed Value: http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | ||
code | 1.. | |||
text | ||||
valueQuantity | .. 0 | |||
valueCode | 1..1 | |||
Slices for coding | 1 .. | Slice: Unordered, Open by value:system | ||
coding:V2Binary | 1..1 | Required V2 Binary code system entry | ||
system | 1.. | Fixed Value: http://terminology.hl7.org/CodeSystem/v2-0136 | ||
code | 1.. | |||
property:quantitiesProperty | The PHG properties represented by quantities such as time synchronization accuracy | |||
type |
The type of time quantity Binding: Quantity 11073 Attributes ( required ) |
|||
coding | 1.. | Required MDC code system entry | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | |||
text | ||||
valueQuantity | 1..1 | |||
value | 1.. | |||
system | 1.. | Fixed Value: http://unitsofmeasure.org | ||
code | 1.. | Fixed Value: us | ||
valueCode | .. 0 | |||
property:codedListProperties | The device properties represented by a list of codes such as the list of certified PHD or H-FS interfaces | |||
type |
What the coded list is Binding: CodeableConcept 11073 Attributes ( required ) |
|||
coding | 1.. | Required MDC code system entry | ||
system | 1.. | Fixed Value: urn:iso:std:iso:11073:10101 | ||
code | 1.. | |||
text | ||||
valueQuantity | .. 0 | |||
valueCode | 1.. | There shall be one valueCode entry for every item supported by the PHG in the list | ||
coding | 1.. | |||
system | 1.. | |||
code | 1.. | |||
Documentation for this format |
Produced 06 Apr 2023